Tsi IAQ10 Manual de usuario Pagina 13

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 20
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 12
11
1. Turn the instrument off. Under the
batteries, move DIP switch # 7 to ON.
Turn the instrument back on.
2. Press and hold the READ key to begin
the calibration procedure. The display
will begin a countdown from 5 to 0.
Release the key when the display reads
0. If the key is released too soon or too
late, the instrument will go back to sam-
pling in real-time. “ZerO” will appear
on the display if the instrument is in
calibration mode.
3. Install the regulator on the zero calibra-
tion gas tank and connect tubing from
the regulator to the inlet fitting on the
top of the Model IAQ10. The regulator
has a fixed flow rate of 0.3 L/min. Do
not use a flow rate lower than
0.25 L/min or higher than 1.0 L/min.
4. Turn on gas flow.
5. Press and release the READ key to begin
a zero reading. The Model IAQ10 begins
a 60-second countdown during the zero
reading. The actual zero reading is
taken in the last 10 seconds. When the
countdown is completed, the display
Vista de pagina 12
1 2 ... 8 9 10 11 12 13 14 15 16 17 18 19 20

Comentarios a estos manuales

Sin comentarios